Advanced Features

| Error Code | Message | Probable Cause | Manual Solution | | :--- | :--- | :--- | :--- | | E001 | Sample Probe Blockage | Clotted blood or fibrin | Run “Probe Clean” cycle. Manually probe with wire tool. | | E010 | Reagent Empty | Diluent or Lyse bottle empty | Replace reagent. Prime the reagent path (Maintenance > Prime). | | E030 | Flow Cell Dirty | Protein buildup | Run 3 cycles of “Flow Cell Clean.” If persists, soak in 10% bleach. | | E100 | HGB Zero Failed | Dirty HGB cell or LED aging | Clean the HGB flow cell. If LED fails, replace (service call). | | WBC Flag | “Turbidity/HB Interf?” | Lipemic or hemolyzed sample | Centrifuge sample and replace plasma with saline. Re-run. | mindray bc 2800 vet user manual

: Internal memory stores up to 20,000 results, including histograms. Connectivity The Essential Guide to the Mindray BC 2800

, such as running a daily calibration or troubleshooting a particular error code? BC-2800VET Troubleshooting Guide | PDF | Home & Garden | | E010 | Reagent Empty | Diluent

: Pre-programmed settings for 13 animal species (including dog, cat, and horse) plus 3 user-defined options. Efficiency